The TriQ is an automated and versatile large chamber triple head sputter coater offering unparalleled ease of use and flexibility based on your application needs. Recommended for controlled thin film deposition for electron microscopy sample preparation and material fabrication.
Features:
- Turbomolecular pumped system with ultimate vacuum of 1 x 10-6 mbar, enabling high quality coatings
- Triple head design offering even deposition of thin film coatings across samples up to 200mm in diameter
- Optimised planetary stages for deposition across 150mm and 200mm diameter samples
- Integrated Film Thickness Monitor (FTM) to control deposition of coating
- Intuitive and responsive touchscreen colour panel design
- Multi-colour LED visual status indicator offering process updates
- Pre-set recipes for standard protocols resulting in sample reproducibility
- Customisable recipes enabling users to tailor process parameters to their specific requirements
- Compatible with a range of target materials
- Sample stages designed to meet a range of application needs are available
Applications:
- Electron Microscopy - High resolution imaging for detailed surface and structural analysis. Enhanced surface conductivity for improved imaging performance. Uniform fine-grained coatings for superior sample preparation. Reduced charging effects and improved analytical accuracy.
- Materials Research - Precision thin film deposition for advanced applications. Controlled materials fabrication with consistent results. Capability to process large-area substrates efficiently. Support for research, development, and prototype production
- Advanced Materials - Novel batteries and fuel cells. Development and study of quantum materials. Semiconductor materials for electronics and microdevices. Photonic materials for optical and communication systems.
Specifications:
|
Standard Stages (Extended or short) |
100mm (4’’) Wafer Stage - Adjustable stage for 2”, 3” and 4” wafers, supplied with a 4” diameter stub holder accommodating up to 18 x 3.2mm pin stubs. 150mm (6’’) Wafer Stage - Adjustable stage for 4” and 6” wafers, supplied with a 6” diameter stub holder accommodating up to 27 x 3.2mm pin stubs. |
|
Stage Options (Extended or short) |
8 Place Stub Stage - Designed for 8 x 25mm or 30mm stubs or resin embedded samples Slide Stage - Microscope slide stage for up to two 75mm x 25mm slides or 8 x 1/8” pin stubs. Tilt Stage - Six stub positions for 15mm, 6.5mm or 3.2mm pin stubs. Tilts up to 90° from horizontal plane. 9 Place Coverslip Stage - Designed to accept 9 x standard coverslips 20mm x 20mm |
| Instrument Size |
440mm (W) x 562mm (H) x 551mm (D), 912mm (H) with chamber open |
| Weight |
52kg (64kg packed) |
| Work chamber |
300mm Ø diameter, 220mm (H) |
| Specimen Stage |
80mm (H) from chamber base (short stage required for sputtering tall samples), 150mm and 200mm Ø diameter substrate. 145mm (H) from chamber base (extended stage required for sputtering short samples), 150mm and 200mm Ø diameter substrate. |
|
Vacuum |
1 x 10-6 mbar (ultimate) 5 x 10-6 mbar(operational) |
|
Vacuum Gauge Range |
1000 mbar to 6.7x10-9 mbar |
| User Interface | Touch screen colour display |
| Supply Voltage |
100-240V, 50-60Hz |
| Power Consumption | 1500VA |
| Deposition Current |
1 - 150mA |
| Deposition Rate |
0.1 – 20nm/min (e.g. for gold) |
| Sputter Timer | 0 - 480 min |
| Sputter Target | Target disc style, 57mm Ø diameter x 0.3mm thickness (gold and chromium supplied as standard) |
|
Film Thickness Monitor (FTM) Quartz Crystal |
Gold, 12mm Ø diameter, 5MHz RC quartz crystals |
| FTM Thickness Range | 1 – 2000nm |
| Site Requirements | Indoor use only Temperature: 15°C to 25°C Humidity: <75% non-condensing Altitude: 2000m Vented gas (preferred): Nitrogen or argon Process gas (preferred): Argon |
TriQ supplied with:
- 3 x Cr (0.5mm thickness x 57mm diameter)
- 100mm and 150mm Wafer Stage (QUO34603)
- Film Thickness Monitor (FTM) Short and Tall (QUO35296)
- Full Range Gauge
- Accessories pack