HI98195 multiparameter waterproof meter
SKU: HI98195 | Availability ⓘ
HI98195 multiparameter waterproof meter
SKU: HI98195 | Availability ⓘ
The HI98195 Multiparameter Waterproof Meter for pH, ORP, EC, TDS, Resistivity, Salinity, Seawater σ and Temperature.
The HI98195 is a waterproof portable logging multiparameter meter that monitors up to 9 different water quality parameters. Its multi-sensor probe allows for the measurement of key parameters including pH, ORP, conductivity, and temperature. The probe transmits readings digitally to the meter, where data points can be displayed and logged. The complete system is simple to setup and easy to use.
Optional shockproof silicon rubber boot - Specially designed to protect your instrument from damage or impact HI710034 Orange 7 Multiparameter 7.34 | w
Features - pH
- Calibration - Up to a three-point calibration with five standard buffers and one custom buffer available
- pH in mV option - Useful for diagnostics
- GLP data - Offset, slope, date, time and buffers used
- Automatically temperature compensated readings
- pH or pH/ORP field replaceable sensors
- Gel filled and maintenance free
- Double junction for reduced contamination of reference cell
Features - EC/TDS/Resistivity
- Calibration - Single-point calibration from six standards
- Temperature compensation - Automatic Temperature Compensation · Configurable temperature coefficient range from 0.00 to 6.00%/°C - Choice of reference temperatures at 20 or 25°C - Absolute conductivity can be displayed along with the temperature compensated value
- Auto-ranging
- Salinity readings - Practical Salinity Scale (PSU) based on conductivity calibration
Specifications
Name | Multiparameter pH/ORP/EC/Pressure/Temperature Waterproof Meter - HI98195 |
pH Range | 0.00 to 14.00 pH |
pH Resolution | 0.01 pH |
pH Accuracy | ±0.02 pH |
pH Calibration | automatic one, two, or three points with automatic recognition of five standard buffers (pH 4.01, 6.86, 7.01, 9.18, 10.01) or one custom buffer |
ORP Range | ±2000.0mV |
ORP Resolution | 0.1mV |
ORP Accuracy | ±1.0mV |
ORP Calibration | automatic at one custom point (relative mV) |
mV Range | ±600.0mV |
mV Resolution | 0.1mV |
mV Accuracy | ±0.5mV |
EC Range | 0 to 9999µS/cm; 0.000 to 200.0mS/cm (absolute EC up to 400.0mS/cm) |
EC Resolution | auto-ranging: 1µS/cm from 0 to 9999µS/cm; 0.01mS/cm from 10.00 to 99.99mS/cm; 0.1mS/cm from 100.0 to 400.0mS/cm; auto-ranging (fixed mS/cm): 0.001mS/cm from 0.000 to 9.999mS/cm; 0.01 from 100.0 to 400.0mS/cm ; manual: 1µS/cm; 0.001mS/cm; 0.01 |
EC Accuracy | ±1% of reading or ±1µS/cm whichever is greater |
EC Calibration | automatic single point, with six standard solutions (84µS/cm, 1413µS/cm, 5.00mS/cm, 12.88mS/cm, 80.0mS/cm, 111.8mS/cm) or custom point |
TDS Range | 0 to 9999ppm (mg/L); 0.000 to 400.0ppt (g/L) (maximum value depends on TDS factor) |
TDS Resolution | manual: 1ppm (mg/L); 0.001ppt (g/L); 0.01ppt (g/L); 0.1 ppt (g/L); 1ppt (g/L); auto-ranging: 1ppm (mg/L) from 0 to 9999ppm (mg/L); 0.01ppt (g/L) from 10.00 to 99.99ppt (g/L); 0. ppt (g/L) from 100.0 to 400.0ppt (g/L); auto-ranging (fixed ppt): |
TDS Accuracy | ±1% of reading or ±1 ppm (mg/L), whichever is greater |
TDS Calibration | based on conductivity calibration |
Salinity (PSU) Range | 0.00 to 70.00 PSU |
Salinity (PSU) Resolution | 0.01 PSU |
Salinity (PSU) Accuracy | ±2% of reading or ±0.01 PSU whichever is greater |
Salinity (PSU) Calibration | based on conductivity calibration |
Salinity (Seawater Sigma) Range | 0.0 to 50.0 σt, σ?, σ?? |
Salinity (Seawater Sigma) Resolution | 0.1 σt, σ?, σ?? |
Salinity (Seawater Sigma) Accuracy | ±1 σt, σ?, σ?? |
Salinity (Seawater Sigma) Calibration | based on conductivity calibration |
Resistivity Range | 0 to 999999 Ω cm; 0 to 1000.0 kΩ cm; 0 to 1.0000 MΩ cm |
Resistivity Resolution | dependent on resistivity reading |
Resistivity Calibration | based on conductivity calibration |
Temperature Range | -5.00 to 55.00°C, 268.15 to 328.15K |
Temperature Resolution | 0.01K, 0.01°C |
Temperature Accuracy | ±0.15°C; ±0.15 K |
Temperature Calibration | automatic at one custom point |
Temperature Compensation | automatic from -5 to 55°C |
Logging Memory | 45,000 records (continuous logging or log-on-demand of all parameters) |
Logging Interval | one second to three hours |
Connectivity | USB to PC with HI9298194 Hanna software installed |
Battery Type/Life | 1.5V AA batteries (4) / approximately 360 hours of continuous use without backlight (50 hours with backlight) |
Environment | 0 to 50°C; RH 100% (IP67) |
Dimensions | 185 x 93 x 35.2mm |
Weight | 400g |