LYNX II automated tissue processors and starter kits
LYNX II automated tissue processors and starter kits
LYNX II is designed to be superior to the Lynx Tissue Processor, with many improvements along with the ability to handle larger size specimens for histology. The Lynx II Automated Tissue Processor for Histology and Microscopy is a cutting-edge tissue processor that can work with any plastic resin or paraffin wax.
The Lynx II has an ability to hold 24 reagent vials for EM processing. HP (Histology processing) is optional and can be achieved with 12 larger size reagent vials. Lynx II has two independently operated heating/cooling stations in both EM and HP modes.
Lynx II operates in batch mode where tissues may be prepared for EM or HP. Both types of tissues are handled in the same way: The sample arm is loaded with samples, and the reagent carousel is loaded with reagents. The Lynx II rotates the reagent carousel depending on the software chosen by the user until the selected reagent station is at the designated location.
Features:
- 12 reagent vials for HP processing, each 125ml volume
- 24 reagent vials for EM processing, each 25ml volume
- 10 Programs for EM and 10 programs for Histology
- Two independently controlled heating/cooling stations
- Easy replacement of reagent carousel
- Paraffin processing
- Sample agitation
- Vacuum infiltration during processing
- Choice of plastic and metal containers for improved chemical compatibility and heat transfer
- Optional external media based interface to a PC for ease of programming and allowing for the editing of programs on a PC and the transfer of programs between the Lynx II and the PC
- Built-in fume extraction and ability to connect the discharge hose to lab exhaust system
- Internal battery back-up during power fail to protect the samples
- Ability to continue run after power interruption
- User Friendly Operator Interface
- Optional external UPS for continued processing during longer power outages
Specifications:
Dimensions: | |
Max Width: | 571.5mm |
Max Depth: | 254mm |
Max Height: | 304.8mm (cover closed) |
Max Weight | 32kg |
Electrical Design | |
Display: | 8 lines x 20 column text only LCD |
Keypad: | Membrane style keypad with tactile feed back |
Mechanism Drive: | Stepper Motors |
Power Requirements: | 230V AC, 50-60Hz, 4A Max |
Back-up power: | NiCd Battery, 12V, 1A-hr |
Electrical: | 5A x 250V AC, Time Delayed (Slo-Blo) |
Operating Environmental Conditions | |
Ambient Temperature: | 20 to 32°C |
Relative Humidity: | 20-80% non-condensing |
Ambient Pressure: | 70-80mm of Hg |
Storage/Transportation Conditions | |
Temperature: | 10-40°C |
Humidity: | 15%-80% non-condensing |
Pressure: | 70-80mm of Hg |
The Lynx II for Microscopy comes with: EM Sample Holder, EM Sample Arm, EM Carousel, Loading Jig, Power Cord, Operator Manual, Aluminium Containers (24), Plastic Containers (72), Plastic Caps (72), Exhaust Hose, Small Baskets (100), Large Baskets (100), Basket Lids (100), 2 Sample Record Note Pads, Spare Fuses.
The Lynx II for Histology comes with: Histo Sample Holder, Histo Sample Arm, Histo Carousel, Power Cord, Operator Manual, Histo Aluminium Containers (24), Histo Plastic Containers (36), Plastic Caps (36), Exhaust Hose, Small Cassettes (100), 2 Sample Record Note Pads, Spare Fuses.
The Lynx II for Microscopy and Histology comes with everything found in both units.
Starter kits available separately:
EMSL12610 |
Histology Kit (Histo Sample Holder, Histo Sample Arm, Histo Carousel, Power Cord, Operator Manual, Histo Aluminium Containers (12), Histo Plastic Containers (36), Plastic Caps (36), Exhaust Hose, Small Cassettes (100), 2 Sample Record Note Pads, Spare Fuses. |
EMSL12612 | Electron Microscopy Kit (EM Sample Holder, EM Sample Arm, EM Carousel, Loading Jig, Power Cord, Operator Manual, Aluminium Containers (12), Plastic Containers (72), Plastic Caps (72), Exhaust Hose, Small Baskets (100), Large Baskets (100), Basket Lids (100), 2 Sample Record Note Pads, Spare Fuses |
Lynx II accessories and spare parts are also available.